# Guava Muse SOP
2+
3+
The Guava Muse Cell Analyzer provides automated cell count and viability measurements. These protocols cover setup, daily operation, and maintenance.
4+
5+
It is highly recommend you read through the manufacturer's protocol prior to using the machine!
6+
7+
- [Cytek-muse-guava-cell-analyzer.pdf](https://github.com/RobertsLab/resources/blob/master/equipment_manuals/Cytek-muse-guava-cell-analyzer.pdf) (GitHub)
8+
9+
## Kits Used
10+
11+
- **Muse System Check Kit**: Verifies the performance of the system by assessing counting accuracy and fluorescence detection using a standardized fluorescent bead reagent. The kit contains a bead reagent and diluent.
12+
- **Muse Count & Viability Kit**: Used to determine viability and total cell count. Accurate assessments can be made with a wide variety of cell lines, even those with unusual culture conditions or a tendency to aggregate.
13+
14+
---
15+
16+
If this is the first time the machine is being _ever_ being used, please see [Instrument Setup Section](#instrument-setup-first-time-only).
17+
18+
## Daily Startup Procedure
19+
20+
1. **Turn On Unit**
21+
- Press power button if not auto-on. Wait for touchscreen to initialize.
22+
23+
2. **Prepare Fluids**
24+
- Empty waste bottle, rinse, and refill with 10 mL bleach.
25+
- Refill cleaning bottle with Guava ICF to indicator line.
26+
- Reconnect fluid tubing.
27+
28+
3. **Reset Fluid Levels**
29+
- On touchscreen: `Menu``System Check``Clean``Reset Fluid Levels`.
30+
31+
4. **Run Complete System Clean**
32+
- On touchscreen: `Menu``Muse System Cleaning``Complete System Clean`.
33+
- Load full tube of Guava Instrument Cleaning Fluid (ICF) → then DI water.
34+
- Run until log confirms success.
35+
36+
5. **Run System Check**
37+
- Mix beads prior to making solution.
38+
- Prepare 1:20 dilution of System Check Beads.
39+
- On touchscreen: `Menu``Essential Tools``System Check`.
40+
- Enter Lot #, Expiration, Check Code (found on bead kit card).
41+
- Mix beads, load tube, run 3 replicates.
42+
- Confirm **PASS** status and %CV ≤10%.
43+
- Export results (optional).
44+
45+
> **Note**: The values recorded during the replicates should be green/grey, meaning they fall within the expected range. If they are red, the values are too high/low, and the system check will fail. This check is temperamental. Values should be within $4.5 \times 10^4$ - $5.5 \times 10^4$.
46+
47+
---
48+
49+
## Daily Shutdown Procedure
50+
51+
1. **Run Complete System Clean**
52+
- Same as startup. Ends with DI water rinse.
53+
54+
2. **Leave DI Water Tube on Loader**
55+
- Never leave bleach or ICF on the system overnight.
56+
57+
3. **Power Down**
58+
- On touchscreen: `Menu``Power Options``Power Off`.
59+
60+
---
61+
62+
## Maintenance Log Example
63+
64+
| Date | User | Waste emptied | ICF refilled | System Clean | System check (p/f) | Notes |
65+
| :--- | :--- | :--- | :--- | :--- | :--- | :--- |
66+
| 18/04/25 | Aakriti | No | Yes | Yes | Pass | First use so did not empty waste |
67+
| 21/04/25 | Aakriti | Yes | Yes | Yes | Pass | |
68+
| 24/04/25 | Aakriti | Yes | Yes | Yes | Pass | |
69+
70+
---
71+
72+
## Important Usage Notes
73+
74+
- **Quick Clean**: When running samples, need to run a quick clean every 20 acquisitions.
75+
76+
--
77+
78+
## Instrument Setup (First-Time Only)
79+
80+
1. **Unpack and Position**
81+
- Carefully remove the Muse from the shipping box and save all packaging.
82+
- Place it on a stable surface away from vibrations.
83+
84+
2. **Connect Fluid Bottles**
85+
- Connect color-coded fluid tubing to the back of the instrument.
86+
- Fill waste bottle: add ~10 mL bleach (to first fill line).
87+
- Fill cleaning solution bottle to the indicator line with Guava ICF.
88+
- Insert bottles into their matching receptacles and connect the tubing.
89+
90+
3. **Insert Flow Cell**
91+
- Open the access door and follow flow cell insertion.
92+
93+
4. **Power On**
94+
- Plug the Muse into a grounded outlet. It turns on automatically the first time.
95+
96+
5. **Log On as Administrator**
97+
- Select Administrator account on first login (p. 35).
98+
- **Important**: Add a new Admin-level user to avoid losing access to data.
99+
100+
6. **Initial Cleaning**
101+
- Run **Complete System Clean** → do this twice on first use (On touchscreen: `Essential Tools` > `Complete System Clean`) (p. 63).

mkdocs.yml

Lines changed: 1 addition & 0 deletions
Original file line numberDiff line numberDiff line change
@@ -87,6 +87,7 @@ nav:
8787
- Lab Protocols:
8888
- Bullet Blender: 'protocols/protocol-bullet_blender.md'
8989
- Glycogen-Glo Assay: 'protocols/protocol-glycogen-glo-assay.md'
90+
- Guava Muse SOP: 'protocols/protocol-guava_muse.md'