HL7 to GPAP bidirectional adapter
This project is a narrow Python MVP that bridges selected HL7 v2 ER7 messages and the Public Invention General Purpose Alarm Protocol (GPAP).
The immediate goal is not to solve all of HL7 alarm interoperability.
The immediate goal is to prove a working bridge that can:
- read a selected HL7 v2 ER7 message
- extract alarm-like data from it
- convert that data into a GPAP alarm string
- optionally publish that GPAP alarm into an ADaM-style MQTT flow
- wrap GPAP alarms and operator responses into custom HL7 Z messages for reverse transport
This repo exists to make the bridge concrete, testable, and easy to evolve.
Public Invention is building an open alarm ecosystem around GPAP, Krake, and ADaM.
HL7 is widely used in healthcare systems.
GPAP is intentionally compact and easy to parse.
A bridge between the two makes it possible to:
- receive alarm-like information from HL7-producing systems
- convert that information into a format usable by ADaM and Krake-style components
- preserve a path back into HL7-shaped messages when needed
This makes the project useful for experimentation in clinical alarm management and other alarm-heavy domains.
This project currently supports a narrow MVP profile.
- Python package with a CLI
- GPAP alarm parsing and rendering
- GPAP operator response parsing and rendering
- HL7 v2 ER7 text parsing with a simple segment splitter
- conversion from a selected HL7 observation message into a GPAP alarm
- conversion from a GPAP alarm into a custom HL7 Z message
- conversion from a GPAP operator response into a custom HL7 Z message
- optional MQTT publish step for ADaM-style integration
- basic tests
- full HL7 interoperability
- FHIR
- complete alarm semantics across all HL7 message types
- full acknowledgment lifecycle with external clinical systems
- production hardening
- robust handling of many local hospital conventions
- multi-OBX alarm correlation
- standardized reverse mapping to an official alarm resource
If you present this repo as a complete HL7 alarm solution, that is false.
This is an MVP bridge.
The current HL7 to GPAP conversion uses a simple, explicit mapping.
The bridge expects HL7 v2 ER7 text and currently treats the first OBX segment as the alarm-carrying observation.
MSH-10→alarm_idMSH-3→sourcePID-3→patient_idOBX-3→ optional GPAPalarm_typeOBX-8→ severityNTE-3→ preferred descriptionOBX-5→ fallback description- default description →
HL7 alarm
N→1L→2LL→3W→3A→4H→4HH→5AA→5
This mapping is heuristic and project-specific.
a<severity>{message_id}[alarm_type]description
Where:
a→ alarmseverity→ digit from 0–5{message_id}→ optional[alarm_type]→ optional- remaining text → description
a5{MSG123}[15074]Pulse exceeded high-high threshold
a→ acknowledges→ shelved→ dismissc→ complete
HL7_to_GPAP/
├── README.md
├── pyproject.toml
├── examples/
│ └── sample_oru.hl7
├── src/
│ └── hl7_to_gpap/
│ ├── __init__.py
│ ├── models.py
│ ├── gpap.py
│ ├── bridge.py
│ ├── mqtt.py
│ └── cli.py
└── tests/
├── test_gpap.py
└── test_bridge.py
py -m pip install -e .If CLI is not on PATH:
py -m hl7_to_gpap.cli --helpCreate examples/sample_oru.hl7:
MSH|^~\&|MONITOR_APP|WARD1|ADAM|PUBINV|20260403113000||ORU^R01|MSG123|P|2.5
PID|1||PAT42^^^HOSP^MR||Doe^Jane
OBR|1||ORDER1|15074^Pulse^LN
OBX|1|NM|15074^Pulse^LN||180|bpm|60-100|HH|||F
NTE|1||Pulse exceeded high-high threshold
py -m hl7_to_gpap.cli hl7-to-gpap --file examples/sample_oru.hl7Output:
a5{MSG123}[15074]Pulse exceeded high-high threshold
py -m hl7_to_gpap.cli gpap-to-hl7 --alarm "a5{MSG123}[15074]Pulse exceeded high-high threshold"Output:
MSH|^~\&|GPAP_BRIDGE|PubInv|HL7_RECEIVER|PubInv|<timestamp>||ZGP^Z01|MSG123|P|2.5
ZAL|MSG123|5|15074|||Pulse exceeded high-high threshold
py -m hl7_to_gpap.cli response-to-hl7 --response "oc{MSG123}|sai|20260403120000"Output:
MSH|^~\&|GPAP_BRIDGE|PubInv|HL7_RECEIVER|PubInv|<timestamp>||ZGP^Z02|MSG123|P|2.5
ZOP|MSG123|c|sai|20260403120000
py -m hl7_to_gpap.cli publish-hl7 \
--file examples/sample_oru.hl7 \
--host public.cloud.shiftr.io \
--port 1883 \
--topic adam/in/alarmsFlow:
- HL7 system emits ER7
- Bridge converts to GPAP
- Bridge publishes to MQTT
- ADaM consumes and processes alarms
- Convert HL7 observations → GPAP alarms
- Feed into ADaM/Krake pipeline
- Market monitoring
- IoT alerting systems
- Keeps downstream systems lightweight
- Moves complexity into the bridge
- No clean standard mapping exists in MVP
- Explicit and honest approach
py -m pytest- GPAP parsing/rendering
- HL7 → GPAP
- GPAP → HL7
- Operator responses
- First OBX only
- Heuristic severity mapping
- No deep HL7 modeling
- Custom reverse mapping
- MQTT is transport-only
- Configurable mappings
- More HL7 support
- Validation improvements
- Integration tests
- Possibly adopt
hl7apy
Early development.
This project proves architecture, not full standards compliance.
AGPL-3.0
The protocol and integration points in that README are grounded in the current GPAP and ADaM repos, including ADaM’s MQTT-based flow and GPAP’s alarm structure with severity, optional message id, and optional alarm type. :contentReference[oaicite:1]{index=1}
Your original README is weak in three ways. It is too vague, it overpromises, and it does not explain the actual bridge behavior. This replacement fixes those problems.